AccuLabAccuLab Life Sciences

GLP or non-GLP: which animal study does your device need?

This page is general information, not regulatory advice for a particular device.

The short answer: FDA's 2023 guidance says that animal studies submitted to support the safety of a medical device must comply with the Good Laboratory Practice regulations, 21 CFR Part 58, though whether they apply to a particular study may depend on its facts and circumstances. Basic exploratory work is outside those regulations, and FDA recommends exploratory animal studies before the safety study in some cases. The rest of this page explains where the line falls.

What GLP covers

Part 58 sets out good laboratory practices for nonclinical laboratory studies that support, or are intended to support, applications to FDA for research or marketing permits, including for medical devices [21 CFR 58.1]. A nonclinical laboratory study is an in vivo or in vitro experiment in which a test article is studied prospectively, under laboratory conditions, to determine its safety [21 CFR 58.3(d)].

Three requirements shape how a GLP study is run. A study director, responsible for the overall conduct of the study, must be designated before the study begins [guidance, Section V; 58.3(m)]. A quality assurance unit, entirely separate from the people who direct and conduct that study, monitors it, inspects it at intervals and reviews the final report against the raw data [58.35]. And the final report, signed by the study director, must include a statement signed by the quality assurance unit giving the dates of its inspections [58.185].

When FDA expects GLP

FDA's guidance on animal studies for medical devices, issued in March 2023, says that animal studies submitted to the agency to support the safety of a medical device must comply with the GLP regulations. It also says it does not try to describe every case in which GLP is required, and that whether the regulations apply to a particular study may depend on its facts and circumstances. FDA recommends that sponsors with specific questions about a study's compliance seek its feedback through the Q-Submission process [FDA guidance, Section II].

What falls outside GLP

Part 58 does not cover basic exploratory studies carried out to find out whether a test article has any potential use, or to determine its physical or chemical characteristics [58.3(d)]. The guidance also leaves such studies outside its own scope [Section III].

FDA recommends exploratory animal studies before the safety study when existing in vivo or in vitro information shows a significant learning curve, for example in placing the device, that would significantly increase the variability of animal responses [Section VI]. It also says the number of animals and the groupings in the GLP study should be designed after exploratory testing, if no other data justify them [Section VI.B].

If a study was not run under GLP

Non-GLP data can still go into a submission, with an explanation. An application for an investigational device exemption must include, for any nonclinical laboratory study it reports, a statement that the study complied with Part 58 or a brief statement of why it did not; failing to comply does not justify leaving out a relevant study [21 CFR 812.27(b)(3)]. A premarket approval application must carry the same statement [21 CFR 814.20(b)(6)(i)]. FDA recommends explaining why, describing all deviations from the regulations, and including information that will help it reconstruct the study [guidance, Section X].

FDA does not certify GLP laboratories

FDA does not license or certify testing facilities for GLP compliance. A facility running a GLP study must meet Part 58 and must let FDA inspect the facility and the study's records and specimens [guidance, Section VII; 21 CFR 58.15].

Before an animal safety study

Before considering an animal study to support a premarket submission, FDA recommends benchtop performance testing of the device in its final finished form. Its guidance suggests considering an animal study when a risk analysis indicates one is needed to assess safety concerns that other methods cannot adequately address [Section IV]. FDA supports the principles of replacing, reducing and refining animal testing, and through the Q-Submission process it will consider whether a proposed alternative method could be assessed as equivalent to an animal test [Section II].

How we work

We support both GLP and non-GLP studies, from early-stage feasibility work to GLP preclinical safety studies, and our scientists and veterinary staff can help design them. A dedicated quality assurance team supports GLP study compliance, and the final report of a GLP study carries a quality assurance statement.

Sources